Standard & high precision probes
HC2A-SM - Stainless steel probe
The HC2A-SM is the robust probe from ROTRONIC for harsh environments and adds to the wide product portfolio. It measures humidity and temperature and calculates the dew/frost point.
Applications
Food, paper, textile, pharmaceutical and cosmetic industries
Features
• Accuracy: ±0.8 %RH, ±0.1 K, at 23 °C ±5 K
• Range of application: -50…100 °C / 0…100 %RH
• Digital interface (UART) and scalable analog outputs, 0…1 V
• Standard output scaling: 0…1 V = -40…60 °C / 0…100 %RH
• Adjusted at 23 °C and 10, 35, 80 %RH
Probe type | Chrome steel standard |
---|---|
Humidity sensor | Hygromer HT-1 |
Temperature sensor | PT100 1/3 DIN Class B |
Long-term stability | <1 %rh / year |
Accuracy at 23 ±5 °C | ±0.8 %rh / ±0.1 K |
Power consumption | 3.3…5 VDC, adjusted at 3.3 VDC |
Dimensions | 15 x 82 mm |
Analogue outputs scaleable by the user | Yes |
Software Update | With internet and HW4 software |
User information | from Software HW4 (devicemanager) |
Data processing via HW4 | with interface cables |
Type of output signals | 2 x 0…1 VDC |
Digital output signal (standard) | UART |
Service Interface | UART |
IP protection rating | IP65 |
FDA/GAMP compatibility | FDA 21 CFR Part 11 and GAMP compatible |
Probe adjustment by software | 1 Point & multipoint %rh & °C |
Device Protection by password | Yes |
Analog output signal (standard) | 0…1 V = 0…100 %rh; 0…1 V = -40…60 °C |
Reverse voltage protection | Mechanically in the connector |
Max. cable length | Up to 5 m with passive cable. Up to 100 m with active cable. |
Weight | 47 g |
Housing | Stainless steel 1.4301 |
Current consumption | 4.5 mA at 3.3 VDC |
Filter type | Wire mesh filter |
Response time | <15 s, without filter |
Maximum wind velocity | 3 m/s, without filter; 25 m/s with wire mesh filter |
Want to learn more about our products?
We will be pleased to personally advise you.
+1 905 754-5164
sales@rotronic.ca