Accredited Calibration Service

NVLAP LAB CODE 20106-0

Rotronic’s calibration laboratory in Hauppauge, New York is accredited by NVLAP to ISO/IEC 17025:2005 and ANSI/NCSL Z540-1-1994 standards for the calibration of humidity and temperature.

The current scope of accreditation is as follows:

HUMIDITY

HygroGen HG2

Relative Humidity Range @ 23C Uncertainty (k=2)
4.5 - 5.5% RH 0.12% RH
9.5 - 10.0% RH 0.15% RH
14.5 - 15.5% RH 0.17% RH
24.0 - 26.0% RH 0.33% RH
34.0 - 36.0% RH 0.39% RH
49.0 - 51.0% RH 0.45% RH
64.0 - 66.0% RH 0.50% RH
79.0 - 81.0% RH 0.55% RH
84.0 - 86.0% RH 0.58% RH



Relative Humidity Range @ 20C Uncertainty (k=2)
89.0 - 91.0% RH 0.65% RH
94.0 - 96.0% RH 0.74% RH


Fixed Humidity Tunnels

Relative Humidity Range @23C
Fixed Humidity Tunnels

Uncertainty (k=2)
0.1…0.8% RH 0.21% RH
10.8…11.8% RH 0.22% RH
34…36% RH 0.29% RH
79…81% RH 0.49% RH



LABORATORY THERMOMETERS

Temperature Measurement Device Calibration

Standard Service

Temperature Range Uncertainty (k=2)
21...27°C 0.065°C



Selected Point Service

Temperature Range Uncertainty (k=2)
10°C 0.09°C
23°C 0.07°C
45°C 0.09°C



Located in a separate, climate controlled area, the environment inside of the calibration laboratory is constantly monitored and maintained. Throughout the calibration process Rotronic’s proprietary software and algorithms ensure stable and reliable calibrations.