Industrial probes HygroClip2 Series

HC2A-IE02-XX Screw-In Probe - Stainless Steel Probe for pressurized conduits and vessels

The Rotronic industrial screw-in probe is especially suitable for high temperatures up to 200 °C and demanding industrial environments with applications in which pressures prevail. It measures humidity and temperature and is compatible with the HF5, HF8 series measurement transmitters, HP32 and HP23 handheld devices and the HL-NT dataloggers.

Advantages of the industrial probe

  • Pressure resistant up to 100 bars
  • Wide temperature measurement range -100…200 °C
  • Outstanding accuracy
  • repeatability and long-term stability
  • Advanced probe housing and construction
  • Hot swappable, no calibration adjustment required due to the digital interface with Rotronic devices

Applications of the probe

  • Industrial manufacturing
  • Drying processes
  • Food and pharmaceutical manufacturing

Smart electronics

  • Based on the Rotronic’s AirChip3000
  • Calculates the dew / frost point
  • Alarm generation
  • Saves adjustment data so that probes can be
  • interchanged without re-adjusting
  • Can be cable mount up to 5 m
  • Hot-swappable

Flexibility and compatibility of the industrial probe

  • User scalable analog ouput signals (2x 0…1V)
  • Digital interface via UART
  • Rapidly interfaced with HygroClip2 compatible devices from Rotronic or for OEM applications
More Information
Dimensions HC2A-IE02-NPT: NPT thread x 104 mm x 2 m Cable
HC2A-IE02-G : G thread x 104 mm x 2 m Cable
Probe type Industrial probe
Humidity sensor Hygromer HT-1
Temperature sensor PT100 1/3 DIN Class B
Long-term stability <1 %rh / year
Measuring range -100...200 °C (short-term peak load) / 0...100 %rh
Accuracy w. adjustment profile "standard" at 23 °C / 10, 35, 80 %rh ± 0.8 %rh / ± 0.1 K
Accuracy at 23 ±5 °C ±0.8 %rh / ±0.1 K
Response time t 63 <15 sec.
Operating range -100…200 °C / 0…100 %rh
Allowed wind speed 3.5 m/s, without filter
Psychrometric calculations Dew or frost point
Analogue outputs scaleable by the user Yes
Software Update With internet and HW4 software
User information from Software HW4 (devicemanager)
Type of output signals 2 x 0…1 VDC
Digital output signal (standard) UART
Service Interface UART
FDA/GAMP compatibility FDA 21 CFR Part 11 and GAMP compatible
Probe adjustment by software 1 Point & multipoint %rh & °C
Analog output signal (standard) 0…1 V = 0…100 %rh; 0…1 V = -40…60 °C
Reverse voltage protection Mechanically in the connector
Max. cable length Up to 5 m with passive cable. Up to 100 m with active cable.
Probe material PPS, Stainless steel 1.4301
Probe protection Filter carrier, brass nickel plated
Filter cartridge depending on the application, order separately
Power supply 3.3…5 VDC
Pressure range 0 – 100 bar

Want to learn more about our products?
We will be pleased to personally advise you.

+41 44 838 11 11
measure@rotronic.ch

GENERAL CONTACT:

Privacy Notice: By submitting the form, you agree that your data will be used to process your request (further information and revocation instructions can be found in the privacy policy).

Copyright © 2021-present Rotronic, Inc. All rights reserved.