Industrial probes HygroClip2 Series
HC2A-IE02-XX Screw-In Probe - Stainless Steel Probe for pressurized conduits and vessels
The Rotronic industrial screw-in probe is especially suitable for high temperatures up to 200 °C and demanding industrial environments with applications in which pressures prevail. It measures humidity and temperature and is compatible with the HF5, HF8 series measurement transmitters, HP32 and HP23 handheld devices and the HL-NT dataloggers.
Advantages of the industrial probe
- Pressure resistant up to 100 bars
- Wide temperature measurement range -100…200 °C
- Outstanding accuracy
- repeatability and long-term stability
- Advanced probe housing and construction
- Hot swappable, no calibration adjustment required due to the digital interface with Rotronic devices
Applications of the probe
- Industrial manufacturing
- Drying processes
- Food and pharmaceutical manufacturing
Smart electronics
- Based on the Rotronic’s AirChip3000
- Calculates the dew / frost point
- Alarm generation
- Saves adjustment data so that probes can be interchanged without re-adjusting
- Can be cable mount up to 5 m
- Hot-swappable
Flexibility and compatibility of the industrial probe
- User scalable analog ouput signals (2x 0…1V)
- Digital interface via UART
- Rapidly interfaced with HygroClip2 compatible devices from Rotronic or for OEM applications
Dimensions | HC2A-IE02-NPT: NPT thread x 104 mm x 2 m Cable HC2A-IE02-G : G thread x 104 mm x 2 m Cable |
---|---|
Probe type | Industrial probe |
Humidity sensor | Hygromer HT-1 |
Temperature sensor | PT100 1/3 DIN Class B |
Long-term stability | <1 %rh / year |
Measuring range | -100...200 °C (short-term peak load) / 0...100 %rh |
Accuracy w. adjustment profile "standard" | at 23 °C / 10, 35, 80 %rh ± 0.8 %rh / ± 0.1 K |
Accuracy at 23 ±5 °C | ±0.8 %rh / ±0.1 K |
Response time t 63 | <15 sec. |
Operating range | -100…200 °C / 0…100 %rh |
Allowed wind speed | 3.5 m/s, without filter |
Psychrometric calculations | Dew or frost point |
Analogue outputs scaleable by the user | Yes |
Software Update | With internet and HW4 software |
User information | from Software HW4 (devicemanager) |
Type of output signals | 2 x 0…1 VDC |
Digital output signal (standard) | UART |
Service Interface | UART |
FDA/GAMP compatibility | FDA 21 CFR Part 11 and GAMP compatible |
Probe adjustment by software | 1 Point & multipoint %rh & °C |
Analog output signal (standard) | 0…1 V = 0…100 %rh; 0…1 V = -40…60 °C |
Reverse voltage protection | Mechanically in the connector |
Max. cable length | Up to 5 m with passive cable. Up to 100 m with active cable. |
Probe material | PPS, Stainless steel 1.4301 |
Probe protection | Filter carrier, brass nickel plated |
Filter cartridge | depending on the application, order separately |
Power supply | 3.3…5 VDC |
Pressure range | 0 – 100 bar |