The Rotronic industrial probe is especially suitable for high temperatures up to 200 °C and demanding industrial environments. It measures humidity and temperature and is compatible with the HF5, HF8 series measurement transmitters, HP32 and HP23 handheld devices and the HL-NT dataloggers.
Advantages of the industrial probe
- Measures relative humidity and temperature
- Wide temperature measurement range -100…200 °C
- Outstanding accuracy
- repeatability and long-term stability
- Advanced probe housing and construction
- Hot swappable, no calibration adjustment required due to the digital interface with Rotronic devices
Applications of the probe
- Industrial Manufacturing
- Drying processes
- Climate chambers
- Brick industry
Smart Electronics
- Based on the Rotronic’s AirChip3000
- Calculates the dew / frost point
- Alarm generation
- Saves adjustment data so that probes can be interchanged without re-adjusting
- Can be cable mount up to 5 m
- Hot-swappable
Flexibility and Compatibility of the industrial probe
- User scalable analog ouput signals (2x 0…1V)
- Digital interface via UART
- Rapidly interfaced with HygroClip2 compatible devices from Rotronic or for OEM applications
Probe type | Industrial probe |
---|---|
Humidity sensor | Hygromer HT-1 |
Temperature sensor | PT100 1/3 DIN Class B |
Long-term stability | <1 %rh / year |
Measuring range | -100...200 °C (short-term peak load) / 0...100 %rh |
Accuracy w. adjustment profile "standard" | at 23 °C / 10, 35, 80 %rh ± 0.8 %rh / ± 0.1 K |
Accuracy at 23 ±5 °C | ±0.8 %rh / ±0.1 K |
Response time t 63 | <15 sec. |
Dimensions | Ø15 x 100 mm / Ø15 x 250 mm / Ø15 x 400 mm / Ø15 x 550 mm / Ø15 x 700 mm |
Operating range | -100…200 °C / 0…100 %rh |
Allowed wind speed | 3.5 m/s, without filter |
Psychrometric calculations | Dew or frost point |
Analogue outputs scaleable by the user | Yes |
Software Update | With internet and HW4 software |
User information | from Software HW4 (devicemanager) |
Type of output signals | 2 x 0…1 VDC |
Digital output signal (standard) | UART |
Service Interface | UART |
FDA/GAMP compatibility | FDA 21 CFR Part 11 and GAMP compatible |
Probe adjustment by software | 1 Point & multipoint %rh & °C |
Analog output signal (standard) | 0…1 V = 0…100 %rh; 0…1 V = -40…60 °C |
Reverse voltage protection | Mechanically in the connector |
Max. cable length | Up to 5 m with passive cable. Up to 100 m with active cable. |
Probe material | PPS |
Probe protection | Filter carrier, brass nickel plated |
Filter cartridge | depending on the application, order separately |
Power supply | 3.3…5 VDC |