HygroFlex8 - HF8 - transmitter
The HF8 is a multi-function transmitter designed to meet the requirements of applications requiring multiple instruments with one self contained instrument. The HF8 has an extensive set of optional functions and settings that can be configured and enabled with the ROTRONIC HW4 software after connecting the HF8 to a PC.
Applications
Measures relative humidity, temperature and all calculated (psychrometric) parameters in the fields of HVAC, industry and pharmaceuticals.
Features
• 2 interchangeable HC2 or analog probes
• Range of application electronics: -40…60 °C / 0…100 %rh
-10…60 °C with LCD
• Digital outputs, also combinable with analog outputs
• Analog inputs (U/I)
• Data logging, up to 10‘000 measured values
• Relay outputs
• Use as simulator for system validation *
• Service interface
Accuracy at 23 ±5 °C | ±0.8 %rh / ±0.1 K |
---|---|
Operating range | -40…85 °C / -10…60 °C (Models with display) 0…100 %rh, non-condensing |
Measuring range with interchangeable probe | 0…100 %rh and -100…200 °C |
Power consumption | 15 ... 40 VDC or 12 ... 28 VAC, 50-60 Hz 9 ... 36 VDC, with galvanic isolation 85 ... 265 VAC, 5 watts, 50-60 Hz with electrical isolation |
Psychrometric calculations | All psychrometric values |
Backlight | with Backlight |
Calibration / adjustment | with Software HW4 |
Display type | Optional graphics display. Presentation configurable. |
Engineering units displayed | 1 or 2 decimal places |
Initialisation time | Typically 1.9 sec. |
Measurement interval | 1 s (typical) |
Number of probe inputs | 2 |
3rd party analogue probe input | 2 |
Keypad (with display only) | Yes |
Alerting | Yes |
Alarm indicators | Yes |
Acoustical alarm (Beeper) | Yes |
Real time clock w.backup battery | Yes |
Memory size | 10,000 pairs of readings (relative humidity and temperature) with HC2; 10,000 readings with 1-channel sensor |
Exploitation software | ROTRONIC HW4 |
Trend indicator | Yes, for models with display |
Probe adjustment by software | Yes, with service cable |
Probe adjustment by keypad | display / keyboard |
Software Update | With internet and HW4 software |
Data processing via HW4 | with interface cables |
Analogue outputs scaleable by the user | Yes |
Analog output signal (standard) | 0…1 V, 0…5 V, 0…10 V, 0(4)…20 mA |
Circuit type | 3/4 wire |
Communication interfaces | RS-485 / USB / LAN |
Configurable relay outputs | 4 |
Relay type and number | SPDT (single pole changeover); Max Switching voltage: 250 VAC*; Max Switching Current: 6A (*resistive load) |
Cable connections | 1x M16x1.5 or ½ "conduit adapter |
Enclosure material | ABS |
Fire protection class | Complies UL94-HB |
IP protection rating | IP65 (except models with USB / Ethernet interface) |
Max. cable length | Up to 5 m with passive cable. Up to 100 m with active cable. |
Max. nr. of devices in RS485 network | 64 |
Max. probe cable length | 10 m |
Probe material | Dependent on probe |
Reverse voltage protection | Protection diode to V + |
Service Interface | Mini USB (UART) |
FDA/GAMP compatibility | FDA 21 CFR Part 11 and GAMP compatible |
Dimensions | 192 x102 x 52 mm |
Weight | 540.0000 |
Want to learn more about our products?
We will be pleased to personally advise you.