HC2A-S3 - Standard meteo probe

HC2A-S3 - Standard meteo probe

The HC2A-S3 probe is the meteorological probe with a white housing - same specification overall as the HC2A-S probe but with a different filter (larger pore size suitable for meteorological applications). It measures relative humidity and temperature and calculates the dew/frost point.

Features of the meteorological probe

  • Accuracy: ±0.8 %rh, ±0.1 K, at 23 °C (10...30 °C)
  • Range of application: -50…100 °C / 0…100 %rh
  • Digital interface (UART) and scalable analog outputs, 0…1 V
  • Standard output scaling: 0…1 V = -40…60 °C / 0…100 %rh
  • Adjusted at 23 °C and 10, 35, 80 %rh

Advantages

  • Measures relative humidity and temperature
  • Outstanding accuracy, repeatability and long-term stability
  • Advanced probe housing and construction
  • Available with interchangeable sensor
  • Hot swappable, no calibration adjustment required due to the digital interface with Rotronic devices

Smart Electronic

  • Based on the Rotronic’s AirChip3000
  • Calculates the dew / frost point
  • Alarm generation
  • Saves adjustment data so that probes can be
  • interchanged without re-adjusting
  • Hot-swappable

Flexibility and Compatibility

  • User scalable analog analog ouput signals (2x 0…1V)1
  • Digital interface via UART2
  • Rapidly interfaced with HygroClip2 devices from
  • Rotronic or in OEM3 applications
More Information
Filter type Polyethylene, 40 μm
Allowed wind speed 20 m/s, with filter
Operating range -50…100 °C / 0…100 %rh
Dimensions Ø15 x 108 mm
Power consumption 3.3...5 VDC, Calibration @ 3.3 VDC / ~4.5 mA
Data processing via HW4 with interface cables
User information from Software HW4 (devicemanager)
Software Update With internet and HW4 software
Analogue outputs scaleable by the user Yes
Psychrometric calculations Dew or frost point
Exploitation software ROTRONIC HW4
Temperature sensor PT100 1/3 DIN Class B
Humidity sensor Hygromer HT-1
Probe type Meteorological probe
Long-term stability <1 %rh / year
Response time t 63 <15 sec.
Reproducibility Airchip <0.02 %rh / 0.01 K
Accuracy w. adjustment profile "High Precision" ±0.5 %RH / ±0.1 K, at 10...30 °C
at 23 °C and 10, 20, 30, 40, 50, 60, 70, 80, 90 %RH
Accuracy w. adjustment profile "standard" ±0.8 %RH / ±0.1 K, at 10...30 °C
Adjustment at 23 °C and 10, 35, 80 %RH
Type of output signals 2 x 0…1 VDC
Digital output signal (standard) UART
Enclosure material Polycarbonate
Filter cartridge Polyethylene dust filter, white, 40 um
Probe protection Polycarbonate plastic cage, white
Max. cable length Up to 5 m with passive cable. Up to 100 m with active cable.
Reverse voltage protection Mechanically in the connector
Weight 17 g
Communication interfaces UART
FDA/GAMP compatibility FDA 21 CFR Part 11 and GAMP compatible
Service Interface UART
Device Protection by password Yes
Analog output signal (standard) 0…1 V = 0…100 %rh; 0…1 V = -40…60 °C
Probe adjustment by software 1 Point & multipoint %rh & °C

Want to learn more about our products?
We will be pleased to personally advise you.

+81 422 502 600
info@michell-japan.co.jp

GENERAL CONTACT:

Datenschutzhinweis: Mit dem Absenden des Formulars erklären Sie sich damit einverstanden, dass Ihre Daten zur Bearbeitung Ihres Anliegens verwendet werden (Weitere Informationen und Widerrufshinweise finden Sie in derDatenschutzhinweis).

Copyright © 2021-present Rotronic, Inc. All rights reserved.