HC2A-S-HH - The Probe for Hydrogen Peroxide Environments.

The Probe for Hydrogen Peroxide Environments (H2O2).

Hydrogen Peroxide Sterilization

Hydrogen peroxide (H2O2) is used to sterilize cleanrooms, incubators and other equipment. In this process the air is saturated by gassing or spraying, resulting in the formation of a microfilm on all surfaces. The hydrogen peroxide kills off all microorganisms. The H2O2 is then either recovered mechanically or allowed to decompose naturally over a period of 48 hours to harmless H2O or O2. The humidity level during the application is crucial for the effectiveness of the process.

Standard humidity sensors have difficulty measuring accurately in high concentration H2O2 environments. The Hydrogen Peroxide inhibits the measurement of water vapour by occupying the pores on the surface of the sensor. This is why ROTRONIC has developed a special sensor, the HYGROMER® HH-1. The life of the sensor is greatly extended compared with that of the standard sensor for these applications. Please see the sensor data sheet for exact technical data on the sensor.

Features

  • Measures relative humidity, temperature and dew point
  • With HYGROMER® HH-1 sensor for hydrogen peroxide environments
  • Highest possible measuring accuracy
  • Conforms to FDA CFR21 Part 11/GAMP
ACCURACY @10…30 °C before and after diffusion phase ±0.8 %rh / ±0.1 K
Dimensions HC2A-S-HH: Ø15 x 109 mm
Peak Load H202 880 ppm / 1200mg/m3
Probe type Hydrogen peroxide Probe
Measuring range 0…60°C / 0...100 %rh
Power supply 3.3…5 VDC
Long-term stability <1 %rh / year
Humidity sensor Hygromer HH-1
Temperature sensor PT100 1/3 DIN Class B
Filter type No filter
Current consumption 4.5 mA at 3.3 VDC
Analogue outputs scaleable by the user Yes
Analog output signal (standard) 0…1 V = 0…100 %rh; 0…1 V = -40…60 °C
Digital output signal (standard) UART
Type of output signals 2 x 0…1 VDC
Max. cable length Up to 5 m with passive cable. Up to 100 m with active cable.
Probe material PPS
Reverse voltage protection Mechanically in the connector
Service Interface UART
FDA/GAMP compatibility FDA 21 CFR Part 11 and GAMP compatible

Want to learn more about our products?
We will be pleased to personally advise you.

+81 422 502 600
info@michell-japan.co.jp

Contact Information

* Required Fields

Privacy Notice: By submitting the form, you agree that your data will be used to process your request (further information and revocation instructions can be found in the privacy policy