Water activity probes

HC2-AW - AW probe

Applications
For water activity measurements in bulk materials such as flour, grain, spices, etc. For solid and pasty products such as meat, sausage as well as oils, fats, etc.

Features
On/Off switch
• UART interface for direct connection to a PC
• Power supply via UART interface
• Adjusted at 23 °C and 10, 35, 80 %rh
•For use with HygroLab C1, HP23-AW-A

More Information
Probe type Measurement probe with UART interface
Humidity sensor HYGROMER® WA-1 / Pt-100, 1/3 DIN Class B
Long-term stability <1 %rh / year
Accuracy at 23 ±5 °C ± 0.008 aw / 0.8 %rh / ±0.1 K
Response time t 63 <20 sec.
Dimensions ø 68 x 60 mm
Psychrometric calculations Dew or frost point
Storage Conditions -40…85 °C
IP protection rating IP20
Connector Standard Rotronic 7-pin connector
AwQuick Mode Yes
Measurement interval 1 s
Communication interfaces UART
Filter cartridge Wire mesh filter with 20…25 μm pore size
Start-up time 1,5 s
CE / EMC conform EMC 2008/108/EG
FDA / GAMP FDA 21 CFR part 11 / GAMP5
Material Stainless steel, aluminum
Power supply Via display instruments
Temperature dependence 0.001 K/K, 0.0002 aw/K
For use with HygroLab C1, HP23-AW-A
T63 90 s

Want to learn more about our products?
We will be pleased to personally advise you.

+65 6376 2107
sg.info@processsensing.com

GENERAL CONTACT:

Privacy Notice: By submitting the form, you agree that your data will be used to process your request (further information and revocation instructions can be found the privacy policy).