Back
HC2-C04 / HC2-C05
Compact Humidity Probe for Tight Spaces
The miniature probe is used for humidity and temperature measurement in confined spaces. It also calculates the dew and frost point and can be mounted discretely.
Applications
Museums, glass cabinets, building material tests, automotive and aviation industries, testing
laboratories, paper, textile and pharmaceutical industries
Features
• Range of application: -40…85 °C / 0…100 %rh
• Digital interface (UART) and scalable analog outputs, 0…1 V
• Standard output scaling: 0…1 V = -40…60 °C / 0…100 %rh
• Adjusted at 23 °C and 10, 35, 80 %rh
Accuracy at 23 ±5 °C | ±1.5 %rh / ±0.3 K |
---|---|
Accuracy w. adjustment profile "standard" | at 23 °C / 10, 35, 80 %rh ± 1.5 %rh / ± 0.3 K |
Probe type | Cable probe |
Operating range | -40…85 °C / 0…100 %rh |
Power consumption | 3.3 VDC ±0.1 / ~4.5 mA |
Long-term stability | <1 %rh / year |
Humidity sensor | Hygromer IN-1 |
Temperature sensor | PT100 1/3 DIN Class B |
Response time t 63 | <15 sec. |
Allowed wind speed | 3 m/s |
Psychrometric calculations | Dew or frost point |
Reproducibility Airchip | <0.02 %rh / 0.01 K |
Exploitation software | ROTRONIC HW4 |
Probe adjustment by software | 1 Point & multipoint %rh & °C |
Software Update | With internet and HW4 software |
User information | from Software HW4 (devicemanager) |
Data processing via HW4 | with interface cables |
Device Protection by password | Yes |
Analogue outputs scaleable by the user | Yes |
Analog output signal (standard) | 0…1 V = 0…100 %rh; 0…1 V = -40…60 °C |
Digital output signal (standard) | UART |
Type of output signals | 2 x 0…1 VDC |
Enclosure material | Polycarbonate |
Probe cable material | PTFE |
Probe material | Stainless steel 1.4305 (V2A) |
Reverse voltage protection | Mechanically in the connector |
Service Interface | UART |
FDA/GAMP compatibility | FDA 21 CFR Part 11 and GAMP compatible |
Want to learn more about our products?
We will be pleased to personally advise you.