Menu

Humidity and Temperature Calibration Services

Rotronic Lifetime Warranty

When using Rotronic's calibration facility in the United States, all probes calibrated at least once every 12 months will receive a lifetime warranty against manufacturing defects.  

Rotronic offers prepaid calibration service plans with significant price advantages.  Be sure to ask your Rotronic Sales Representative about our prepaid calibration plans


Accredited Calibration Service

NVLAP LAB CODE 20106-0

Rotronic’s calibration laboratory in Hauppauge, New York is accredited by NVLAP to ISO/IEC 17025:2005 and ANSI/NCSL Z540-1-1994 standards for the calibration of humidity and temperature.

The current scope of accreditation is as follows:

HUMIDITY

HygroGen HG2

Relative Humidity Range @ 23C Uncertainty (k=2)
4.5 - 5.5% RH 0.12% RH
9.5 - 10.0% RH 0.15% RH
14.5 - 15.5% RH 0.17% RH
24.0 - 26.0% RH 0.33% RH
34.0 - 36.0% RH 0.39% RH
49.0 - 51.0% RH 0.45% RH
64.0 - 66.0% RH 0.50% RH
79.0 - 81.0% RH 0.55% RH
84.0 - 86.0% RH 0.58% RH



Relative Humidity Range @ 20C Uncertainty (k=2)
89.0 - 91.0% RH 0.65% RH
94.0 - 96.0% RH 0.74% RH


Fixed Humidity Tunnels

Relative Humidity Range @23C
Fixed Humidity Tunnels

Uncertainty (k=2)
0.1…0.8% RH 0.21% RH
10.8…11.8% RH 0.22% RH
34…36% RH 0.29% RH
79…81% RH 0.49% RH



LABORATORY THERMOMETERS

Temperature Measurement Device Calibration

Standard Service

Temperature Range Uncertainty (k=2)
21...27°C 0.065°C



Selected Point Service

Temperature Range Uncertainty (k=2)
10°C 0.09°C
23°C 0.07°C
45°C 0.09°C



Located in a separate, climate controlled area, the environment inside of the calibration laboratory is constantly monitored and maintained. Throughout the calibration process Rotronic’s proprietary software and algorithms ensure stable and reliable calibrations.



Documentation