Menu
Back

RMS CCA-S-20X-SET

The CCA-S-20X is an analog probe with NDIR technology for measurement of carbon dioxide (CO2) for incubators. Developed with a pyroelectric infrared detector with dual temperature compensation and an integrated semiconductor temperature sensor to maximize accuracy and minimize drift. The signal is converted to 4...20 mA by the converter with power supply (CCA-S-20X-SET).

Benefits

  • Measures from 0 to 20 %CO2
  • Suitable for incubators: 37 °C, 95…98 %RH and 5 %CO2
  • Interchangeable probes
  • ±0.1 °C accuracy in measurement range

Applications

  • Incubators

Delivery Package:

  • CCA-S-20X
  • Power supply
  • Converter box
  • E2-01XX

Compatible with:

  • RMS-MADC-868/915-A, Analog Input
  • CCA-S-20X CO2, Sensor
  • RMS-WEB, RMS Server Software
  • RMS-CLD, RMS Cloud
Accuracy ±10 % of measured value
Dimensions Sensor: 32 mm x 87 mm
Converter box: 100 mm x 77 mm x 40 mm
Parameters CO2 concentration (ppm / %)
Range of application -20…50 °C / 0…100 %RH, 700…1200 hPa
Measurement principle Infrared (NDIR)
Measurement range 0…20 %CO2
Pressure dependence ±0.15 % of measured value/hPa
Medium Air and non-aggressive gases
Power supply CCA-S-20X: 3...5 VDC / 80 mA
CCA-S-20X-Set: 24 VDC / 150 mA
AC adapter requirements 100…240 VAC / 50…60 Hz / 0.3 A
Long-term stability ±0.24 %CO2/year
Temperature dependence ±10 % of measured value
Start-up time 60s
Enclosure material Polycarbonate (housing)
Stainless steel DIN 1.4305 (nut)
Fire protection class Complies UL94-HB
FDA / GAMP FDA 21 CFR part 11 / GAMP5
Output signals 4...20 mA (CCA-S-20X-SET)
0.4...2.0 V (CCA-S-20X only sensor)
Storage and transport conditions -20...30 °C / 0...95 %RH
IP protection class IP40
Weight 55 g sensor
200 g converter box

Want to learn more about our products?
We will be pleased to personally advise you.

+41 44 838 11 11