CO2 – Measurement and Monitoring of Carbon Dioxide in Rooms

CP11 - Handheld Instrument for CO2, Humidity and Temperature

The CP11 handheld instrument is an economical multiple parameter meter that simultaneously measures and records CO2, humidity and temperature. It also calculates the dew point and wet bulb temperature. Equipped with the field-tested ROTRONIC HYGROMER® IN-1 humidity sensor, this instruments offers unbeatable value for money. Using the ROTRONIC SW21 software that comes with the device, it can be set as required and data can be downloaded, saved and analyzed.

Applications
Indoor air quality, monitoring in underground car parks (safety), Leak monitoring in filling plants (safety).

Features
• Measures and logs CO2, relative humidity and temperature
• Calculates the dew point
• Hygromer® IN-1 humidity sensor
• 18,000 data point memory for CO2, humidity and temperature values
• Maximum, minimum and average value display
• Adjustable audible CO2 alarm
• Display with backlight
• Adjustable auto power off function
• Mini USB interface

Software
Supplied with the instrument or available as download SW2.1 (settings, data download, display of measured values) or compatible with optionally available HW4 software.

More Information
Humidity sensor ROTRONIC HYGROMER® IN-1
Temperature sensor Thermistor
Long-term stability < 1.5 %RH per year
Measuring range 0…5000 ppm, 0.1...99.95 %RH, -20...60 °C
Accuracy at 23 ±5 °C ±30 ppm ±5 % of the measured value, ±2 %rF (10...90 %rF), others ± 5 %rF, ±0,3 K
Response time t 63 <30 s, without filter
Alarm functions Adjustable for CO2 measurement
Dimensions 76,7 x 41,8 x 233,9 mm
IP protection rating IP30
Weight 290 g
Housing ABS
Parameters shown CO2, relative humidity and temperature
Calculations Dew point, wet bulb temperature
Clock Real time clock
Current consumption 40 mA (typical)
CE / EMC conform 2004/108/EC
Power supply 4 x AA batteries / optional AC adapter AC1214
Battery lifetime 50 h
Measured value memory 18,000 values with time stamp, automatic recording (%RH / °C / ppm), 99 single values with time stamp, manual recording, (%RH / °C / DP / Tw / ppm)
Measurement principle Non dispersive infrared (NDIR) with automatic baseline correction (ABC)
Response time <10 sec @ 30 cc/min. flow, <3 min diffusion time
Adjustment points 0, 400 ppm and 35, 80 %RH
Pressure dependence +1.6 % reading per kPa
Zero drift <10 ppm/year
Maintenance No maintenance (standard indoor application)
Warm-up time <1 min.

Want to learn more about our products?
We will be pleased to personally advise you.

+41 44 838 11 11
measure@rotronic.ch

GENERAL CONTACT:

Privacy Notice:By submitting the form, you agree that your data will be used to process your request (further information and revocation instructions can be found in the privacy policy).

Copyright © 2021-present Rotronic, Inc. All rights reserved.