Back
RMS CCA-S-20X-SET
The CCA-S-20X is an analog probe with NDIR technology for measurement of carbon dioxide (CO2) for incubators. Developed with a pyroelectric infrared detector with dual temperature compensation and an integrated semiconductor temperature sensor to maximize accuracy and minimize drift. The signal is converted to 4...20 mA by the converter with power supply (CCA-S-20X-SET).
Benefits
- Measures from 0 to 20 %CO2
- Suitable for incubators: 37 °C, 95…98 %RH and 5 %CO2
- Interchangeable probes
- ±0.1 °C accuracy in measurement range
Applications
- Incubators
Delivery Package:
- CCA-S-20X
- Power supply
- Converter box
- E2-01XX
Compatible with:
- RMS-MADC-868/915-A, Analog Input
- CCA-S-20X CO2, Sensor
- RMS-WEB, RMS Server Software
- RMS-CLD, RMS Cloud
Accuracy | ±10 % of measured value |
---|---|
Dimensions | Sensor: 32 mm x 87 mm Converter box: 100 mm x 77 mm x 40 mm |
Parameters | CO2 concentration (ppm / %) |
Range of application | -20…50 °C / 0…100 %RH, 700…1200 hPa |
Measurement principle | Infrared (NDIR) |
Measurement range | 0…20 %CO2 |
Pressure dependence | ±0.15 % of measured value/hPa |
Medium | Air and non-aggressive gases |
Power supply | CCA-S-20X: 3...5 VDC / 80 mA CCA-S-20X-Set: 24 VDC / 150 mA |
AC adapter requirements | 100…240 VAC / 50…60 Hz / 0.3 A |
Long-term stability | ±0.24 %CO2/year |
Temperature dependence | ±10 % of measured value |
Start-up time | 60s |
Enclosure material | Polycarbonate (housing) Stainless steel DIN 1.4305 (nut) |
Fire protection class | Complies UL94-HB |
FDA / GAMP | FDA 21 CFR part 11 / GAMP5 |
Output signals | 4...20 mA (CCA-S-20X-SET) 0.4...2.0 V (CCA-S-20X only sensor) |
Storage and transport conditions | -20...30 °C / 0...95 %RH |
IP protection class | IP40 |
Weight | 55 g sensor 200 g converter box |
Want to learn more about our products?