RMS CCA-S-20X-SET (EOL, Replacement product on request)
RMS CCA-S-20X-SET (EOL, Replacement product on request)
The CCA-S-20X is an analog probe with NDIR technology for measurement of carbon dioxide (CO2) for incubators. Developed with a pyroelectric infrared detector with dual temperature compensation and an integrated semiconductor temperature sensor to maximize accuracy and minimize drift. The signal is converted to 4...20 mA by the converter with power supply (CCA-S-20X-SET).
Benefits
- Measures from 0 to 20 %CO2
- Suitable for incubators: 37 °C, 95…98 %RH and 5 %CO2
- Interchangeable probes
- ±0.1 °C accuracy in measurement range
Applications
- Incubators
Delivery Package:
- CCA-S-20X
- Power supply
- Converter box
- E2-01XX
Compatible with:
- RMS-MADC-868/915-A, Analog Input
- CCA-S-20X CO2, Sensor
- RMS-WEB, RMS Server Software
- RMS-CLD, RMS Cloud
| 尺寸 | Sensor: 32 mm x 87 mm Converter box: 100 mm x 77 mm x 40 mm |
|---|---|
| 长期稳定性 | ±0.24 %CO2/year |
| Connector | Standard Rotronic 7-pin connector |
| 阻燃类 | 符合UL94-HB |
| 外壳材质 | Polycarbonate (housing) Stainless steel DIN 1.4305 (nut) |
| 重量 | 55 g sensor 200 g converter box |
| 启动时间 | 60s |
| FDA / GAMP | FDA 21 CFR part 11 / GAMP5 |
| 供电 | CCA-S-20X: 3...5 VDC / 80 mA CCA-S-20X-Set: 12 VDC / 0,5 A |
| 温度分辨率 | ±10 % of measured value |
| 输出信号 | 4...20 mA (CCA-S-20X-SET) 0.4...2.0 V (CCA-S-20X only sensor) |
| T63 | 90 s |
| 测量原理 | 红外线(NDIR) |
| Pressure dependence | ±0.15 % of measured value/hPa |
| 参数 | CO2 concentration (ppm / %) |
| 精度 | ±0.2 %CO2 |
| 中等 | 空气和非侵蚀性气体 |
| 测量范围 | 0…20 %CO2 |
| AC adapter requirements | 100…240 VAC / 50…60 Hz / 0.3 A |
| IP protection class | IP40 |
Want to learn more about our products?
We will be pleased to personally advise you.
+86 21 5401 2255
info.cn@processsensing.com