RMS CCA-S-20X-SET

RMS CCA-S-20X-SET

The CCA-S-20X is an analog probe with NDIR technology for measurement of carbon dioxide (CO2) for incubators. Developed with a pyroelectric infrared detector with dual temperature compensation and an integrated semiconductor temperature sensor to maximize accuracy and minimize drift. The signal is converted to 4...20 mA by the converter with power supply (CCA-S-20X-SET).

Benefits

  • Measures from 0 to 20 %CO2
  • Suitable for incubators: 37 °C, 95…98 %RH and 5 %CO2
  • Interchangeable probes
  • ±0.1 °C accuracy in measurement range

Applications

  • Incubators

Delivery Package:

  • CCA-S-20X
  • Power supply
  • Converter box
  • E2-01XX

Compatible with:

  • RMS-MADC-868/915-A, Analog Input
  • CCA-S-20X CO2, Sensor
  • RMS-WEB, RMS Server Software
  • RMS-CLD, RMS Cloud
More Information
Dimensions Sensor: 32 mm x 87 mm
Converter box: 100 mm x 77 mm x 40 mm
Long-term stability ±0.24 %CO2/year
Fire protection class Complies UL94-HB
Enclosure material Polycarbonate (housing)
Stainless steel DIN 1.4305 (nut)
Weight 55 g sensor
200 g converter box
Start-up time 60s
FDA / GAMP FDA 21 CFR part 11 / GAMP5
Power supply CCA-S-20X: 3...5 VDC / 80 mA
CCA-S-20X-Set: 12 VDC / 0,5 A
Temperature dependence ±10 % of measured value
Output signals 4...20 mA (CCA-S-20X-SET)
0.4...2.0 V (CCA-S-20X only sensor)
Measurement principle Infrared (NDIR)
Pressure dependence ±0.15 % of measured value/hPa
Parameters CO2 concentration (ppm / %)
Accuracy ±0.2 %CO2
Medium Air and non-aggressive gases
Measurement range 0…20 %CO2
AC adapter requirements 100…240 VAC / 50…60 Hz / 0.3 A
IP protection class IP40