HygroClip XD - Standard probe
HygroClip XD - Standard probe
Versatile HygroClip2 - Standard probe
Features
• Measures relative humidity & temperature
• Calculates dew/frost point
• Power supply range: 5…16 VAC / 5…24 VDC
• Outputs: 0...1 V / 0...5 V / 0...10 V
• Voltage outputs configurable via software
• Highest precision and reproducibility
• Excellent long-term stability <1 %RH / year
• Digital interface for calibration and measurement processing
| Humidity sensor | ROTRONIC HYGROMER® IN-1 |
|---|---|
| Temperature sensor | PT100 1/3 DIN Class B |
| Long-term stability | <1 %rh / year |
| Accuracy at 23 ±5 °C | ±0.8 %rh / ±0.2 K |
| Response time t 63 | <15 sec. |
| Power consumption | 5…16 VAC / 5…24 VDC |
| Dimensions | 135 mm, ∅ 15 mm |
| Operating range | -40..85 °C / 0…100 %rF |
| Psychrometric calculations | Dew and frost point |
| Digital output signal (standard) | UART or RS-485 |
| IP protection rating | IP65 |
| Probe adjustment by software | Single point, multipoint, dew point (%RH / °C), with service cable |
| Analog output signal (standard) | 0…1 / 5 / 10 V |
| Max. probe cable length | 2 m / 5 m |
| Probe protection | Polycarbonate plastic cage |
| Filter cartridge | Polyethylene dust filter, gray, 20 um |
| Housing | Polycarbonate |
| Current consumption | 7 mA (typical) @ 24 VDC, 0…10 V / 12 mA (typical) @ 5 VDC, 0…1 V |
| CE / EMC conform | 2007 / 108 / EG |
| FDA / GAMP | FDA 21 CFR part 11 / GAMP5 |
| Maintenance | Annual calibration recommended |
| Maximum wind velocity | 20 m/s, with polyethylene filter |
Want to learn more about our products?
We will be pleased to personally advise you.
+65 6376 2107
sg.info@processsensing.com